恒溫烙鐵常見7大故障維修方法
1、發熱指示燈不亮
依次(ci)檢查(cha)電線松脫,保險絲,24V,烙鐵短(duan)路,接地彈簧(huang)。
檢查保險(xian)時,先關電源,拔下(xia)烙鐵,將恒溫臺翻過來,擰松螺絲(si),再(zai)反過來,將上(shang)蓋輕取下(xia)來。
如果(guo)24V沒(mei)問題,則檢查(cha)烙(luo)鐵是否(fou)短路(lu)。
烙鐵頭(tou)只能用清(qing)(qing)潔海綿清(qing)(qing)洗,不可用紗紙打(da)磨(mo)。
2、燈亮烙鐵不升溫
依次(ci)檢查電線破損,發熱元件損壞。
萬用表調到蜂鳴(ming)檔,測量(liang)線(xian)(xian)(xian)(xian)是否破損方(fang)法,測量(liang)烙(luo)鐵頭(tou)針(zhen)頭(tou)與線(xian)(xian)(xian)(xian)連接情況,逆時(shi)針(zhen)1、2、3、4、5,1為(wei)紅線(xian)(xian)(xian)(xian),只要響了,說(shuo)明相通;2為(wei)白線(xian)(xian)(xian)(xian),3為(wei)地線(xian)(xian)(xian)(xian),4為(wei)綠線(xian)(xian)(xian)(xian),5為(wei)藍線(xian)(xian)(xian)(xian)。如果有(you)不相通線(xian)(xian)(xian)(xian),則更換線(xian)(xian)(xian)(xian)。
如(ru)果(guo)線(xian)(xian)(xian)沒問題,則檢查發(fa)(fa)熱(re)元件(jian)。4、5分(fen)別是綠線(xian)(xian)(xian)和藍線(xian)(xian)(xian),它們連接的(de)是發(fa)(fa)熱(re)器(qi),正(zheng)(zheng)常(chang)阻值在(zai)2.6-3.8之(zhi)間(jian),如(ru)果(guo)異(yi)(yi)常(chang),需(xu)更換(huan)陶瓷(ci)發(fa)(fa)熱(re)器(qi)。1、2是紅線(xian)(xian)(xian)和白線(xian)(xian)(xian),它們連的(de)是傳感器(qi),正(zheng)(zheng)常(chang)阻值在(zai)43-58之(zhi)間(jian),如(ru)果(guo)異(yi)(yi)常(chang)需(xu)更換(huan)傳感器(qi)。
3、烙鐵斷斷續續升溫
根據上(shang)邊方法檢查(cha)電線是否破損。
4、烙鐵頭不沾錫
依次檢查是否溫度(du)太(tai)高(gao),清洗烙鐵頭。
溫度太高,則需降低溫度。如果烙鐵頭氧化,則清洗烙鐵頭。電烙鐵頭不沾錫處理辦法>>
5、烙鐵頭溫度太低
依次檢查烙鐵頭是(shi)否(fou)氧化,校準溫度。
烙鐵頭溫(wen)度(du)太低(di),也(ye)是先清洗烙鐵頭。如(ru)果更換過元器件(jian),則需校準溫(wen)度(du)(用螺絲刀擰旋轉(zhuan)下邊(bian)的小螺絲)。
6、烙鐵頭拆不開
檢查烙鐵頭是否(fou)生銹,如果(guo)是,將烙(luo)鐵頭(tou)和發熱元件一(yi)起更換。
7、烙鐵頭達不到所需溫度
也是校準溫度。